In April 2016 Manchester eScholar was replaced by the University of Manchester’s new Research Information Management System, Pure. In the autumn the University’s research outputs will be available to search and browse via a new Research Portal. Until then the University’s full publication record can be accessed via a temporary portal and the old eScholar content is available to search and browse via this archive.

Modelling and Correction of Scatter in a Switched Source Multi-Ring Detector X-ray CT machine

Wadeson, Nicola Lisa

[Thesis]. Manchester, UK: The University of Manchester; 2011.

Access to files

Abstract

The RTT80 cone beam x-ray computed tomography system, developed by Rapiscan Systems Ltd, uses switched x-ray sources and fixed offset detector rings to remove the time consuming mechanical rotations of earlier imaging systems. This system produces three-dimensional images in real time. A Geant4 Monte Carlo simulation has been developed to investigate scattered radiation in the uncollimated detector machine, showing high levels of scatter behind highly attenuating objects.A new scatter correction method is proposed which estimates scatter to each detector, in each projection, from 1cm$^{3}$ voxels of the computerised object. The scatter distributions from different materials are pre-determined using a Geant4 Monte Carlo simulation. The intensity of scatter from each voxel is based on measured data. The method is applied to two simulated test objects, a water box simulated with a monoenergetic input spectrum and a test suitcase simulated with a polyenergetic spectrum. The test suitcase is broken down into separate components to analyse the method further. The results show that the method performs well for low attenuating objects, but the results are sensitive to the intensity values. However, the method provides a good basis for a scatter correction method.

Bibliographic metadata

Type of resource:
Content type:
Form of thesis:
Type of submission:
Degree type:
Doctor of Philosophy
Degree programme:
PhD Mathematical Sciences
Publication date:
Location:
Manchester, UK
Total pages:
138
Abstract:
The RTT80 cone beam x-ray computed tomography system, developed by Rapiscan Systems Ltd, uses switched x-ray sources and fixed offset detector rings to remove the time consuming mechanical rotations of earlier imaging systems. This system produces three-dimensional images in real time. A Geant4 Monte Carlo simulation has been developed to investigate scattered radiation in the uncollimated detector machine, showing high levels of scatter behind highly attenuating objects.A new scatter correction method is proposed which estimates scatter to each detector, in each projection, from 1cm$^{3}$ voxels of the computerised object. The scatter distributions from different materials are pre-determined using a Geant4 Monte Carlo simulation. The intensity of scatter from each voxel is based on measured data. The method is applied to two simulated test objects, a water box simulated with a monoenergetic input spectrum and a test suitcase simulated with a polyenergetic spectrum. The test suitcase is broken down into separate components to analyse the method further. The results show that the method performs well for low attenuating objects, but the results are sensitive to the intensity values. However, the method provides a good basis for a scatter correction method.
Thesis main supervisor(s):
Funder(s):
Language:
en

Institutional metadata

University researcher(s):

Record metadata

Manchester eScholar ID:
uk-ac-man-scw:125710
Created by:
Wadeson, Nicola
Created:
30th June, 2011, 07:49:58
Last modified by:
Wadeson, Nicola
Last modified:
1st May, 2015, 19:22:29

Can we help?

The library chat service will be available from 11am-3pm Monday to Friday (excluding Bank Holidays). You can also email your enquiry to us.