In April 2016 Manchester eScholar was replaced by the University of Manchester’s new Research Information Management System, Pure. In the autumn the University’s research outputs will be available to search and browse via a new Research Portal. Until then the University’s full publication record can be accessed via a temporary portal and the old eScholar content is available to search and browse via this archive.

A comparison of inter-aural attenuation with the Etymotic ER-3A insert earphone and the Telephonics TDH-39 supra-aural earphone.

Munro, K J; Agnew, N

British journal of audiology. 1999;33(4):259-62.

Access to files

Full-text and supplementary files are not available from Manchester eScholar. Full-text is available externally using the following links:

Full-text held externally

Abstract

One of the many reported advantages of the insert earphone over the supra-aural earphone is increased inter-aural attenuation (IA). Minimum values of IA determine the need for masking of the non-test ear in air-conduction audiometry. The aim of the present study was to measure inter-aural attenuation for the Etymotic Research ER-3A insert earphone (with deep and shallow insertion of the ear plug within the ear canal) and compare this with the supra-aural Telephonics TDH-39/MX41-AR earphone/cushion combination. Subjects were 18 adults ranging in age from 38 to 68 years (mean 50 years). Each subject had no hearing in one ear following translabyrinthine surgery for removal of an acoustic neuroma. The opposite ear had hearing thresholds better than 40 dB HL and an air-bone gap of less than 10 dB at any audiometric frequency. Pure tone air-conduction thresholds were obtained in the range 0.25-8 kHz. Deep insertion of the insert earphone was deemed to occur when the outside edge of the ear plug was flush with the entrance of the ear canal. Shallow insertion was deemed to occur when half of the ear plug (6 mm) was inside the entrance of the ear canal. IA was defined operationally as the difference between the good-ear and poor-ear not-masked air conduction threshold for a given audiometric frequency and earphone. The results show that the TDH-39/MX41-AR combination provides a median IA of approximately 60 dB with a lower limit of approximately 45 dB. Greater IA was obtained with the ER-3A insert earphone but this depended on the depth of insertion. With a deep insertion, the 1A values were some 15-20 dB greater than with the supra-aural earphone. Although frequency-specific IA values are provided, a simple rule of thumb is to apply masking to the non-test ear when the pure tone airconduction signal from the ER-3A insert earphone exceeds the bone conduction threshold of the non-test ear by 55 dB HL or more. If it is not possible to obtain a deep insertion depth this value should be reduced by 5 dB.

Bibliographic metadata

Type of resource:
Content type:
Publication type:
Author(s):
Published date:
Abbreviated journal title:
ISSN:
Place of publication:
ENGLAND
Volume:
33
Issue:
4
Pagination:
259-62
Pubmed Identifier:
10509860
Access state:
Active

Institutional metadata

University researcher(s):

Record metadata

Manchester eScholar ID:
uk-ac-man-scw:146784
Created by:
Munro, Kevin
Created:
10th January, 2012, 12:15:37
Last modified by:
Munro, Kevin
Last modified:
10th January, 2012, 12:15:37

Can we help?

The library chat service will be available from 11am-3pm Monday to Friday (excluding Bank Holidays). You can also email your enquiry to us.